Hardness Calcium (B) T / M191
Hardness Calcium (B) T
20 - 500 mg/l CaCO
Murexide
Material
EN
Required material (partly optional):
Reagents
Set Calcio H No. 1/No. 2 100 Pc.
Set Calcio H No. 1/No. 2 250 Pc.
Preparation
1. Strong alkaline or acidic water samples should be adjusted between pH 4 and pH 10
before the analysis (use 1 mol/l Sulphuric acid or 1 mol/l Sodium hydroxide).
Notes
1. To optimise the readings, an optional batch-specific blind value method can be
performed (see photometer description).
2. For accurate results, exactly 10 ml of water sample must be used for the test.
3. This method was developed from a volumetric procedure. Due to undefined boundary
conditions, deviations from the standardised method may be greater.
4. The method works in the high measuring range with greater tolerances than in the
low measuring range. When diluting samples, always measure in the first third of the
range.
50
3
#
#
EN Method Reference Book 1.0
Packaging Unit
Part Number
100 each
517761BT
250 each
517762BT
M191
CAH