Appendix II EMC Guidance and Manufacturer Declaration
Table 1:
Guidance and manufacturer's declaration –electromagnetic emission
The Infrared Thermometer is intended for use in the electromagnetic environment specified
below. The purchaser or the user of the device should assure that it is used in such
environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Table 2:
Guidance and manufacturer's declaration-electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic environment specified
below. The purchaser or the user of the Infrared Thermometer should assure that it is used in
such environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage vatiations on power
supply input lines
IEC 61000-4-11
Power frequency (50 / 60Hz)
magnetic field
IEC 61000-4-8
Compliance
Group 1
Class A
Class A
Not applicable
IEC60601
test level
±8kV contact
± 15 kV air
±2kV for power supply
lines
± 1 kV for input/output line
±1 kV lines to lines
±2 kV lines to earth
<5%UT(>95%dip in UT)
for 0.5 cycle
40% UT(60%dip in UT) for
5 cycle
70%UT(30%dip in UT) for
25 cycle
<5%UT(>95%dip in UT)
for 5 sec
30 A/m
64
Compliance level
±8kV contact
±15kV air
±2kV for power supply lines
Not Applicable
±1 kV lines to lines
±2 kV lines to earth
<5%UT(>95%dip in UT) for
0.5 cycle
40% UT(60%dip in UT) for
5 cycle
70%UT(30%dip in UT) for
25 cycle
<5%UT(>95%dip in UT) for
5 sec
30A/m