Rhythm 2.
Gain (sensitivity): use the
mm/mV and 20 mm/mV. The overall gain (sensitivity) can be checked by calibration function.
Speed: use the
mm/s.
Filter: use the
AC+EMG, AC+DFT, EMG+DFT and AC+EMG+DFT.
In which, AC
EMG
DFT
Display calibration signal: after pressing
screen once.
Note: The calibration is an automatic process, user does not need to press any button.
Print/End print: use the
Auto mode: After starting to print, the system automatically prints and stores the real-time
12-lead ECG waveform. The length is determined by the relevant settings in the print setup.
Based on the settings, the automatic analysis data and conclusions are printed, and the system
automatically ends printing.
Manual mode: After starting to print, user need to switch the lead to print the waveform of
different leads, that is, the ECG printed in the manual mode is asynchronous, and the data is not
saved. User need to press the PRINT button again when the print needs to be terminated.
During printing, the display content of print status are including:
Display content
Process...
Waiting...
No Paper.
Print Timeout.
ECG Timeout
Low Power
Note: You can not print until ECG waveforms are displayed on the screen.
button to switch the gain between 5 mm/mV, 10
button to switch the speed between 12.5 mm/s, 25 mm/s and 50
button to switch the filter between no filter, AC, EMG, DFT,
AC filter
EMG filter
Baseline filter
button to start or end the printing operation.
Explanation
It is printing.
It is finishing printing.
Lack of paper, user should restart printing after loading paper.
Communication failure between this system and printing
sub-system.
Communication failure between this system and sampling
sub-system.
Low power, it cannot start printing.
button,1 mV signal will occur on the