Reassembling your mask
1. Align and insert the frame connector tab (a) into the cushion slot (b)
and press down until it clicks. Repeat on the other side.
2. Attach the elbow to the top of the frame until it clicks.
3. With the ResMed logo facing outwards and up, insert the upper
headgear straps into the frame from the inside and fold the fastening
tabs over to secure.
Note: If the elbow ring detaches, re-insert into the top of the frame.
WARNING
• Magnets are used in the lower headgear straps and the frame of the mask. Ensure the
headgear and frame is kept at least 50 mm away from any active medical implant (eg,
pacemaker or defibrillator) to avoid possible effects from localized magnetic fields.
The magnetic field strength is less than 400 mT.
• The mask contains safety features, the exhaust vent holes and anti-asphyxia valves, to
enable normal breathing and exhaust exhaled breath. Occlusion of the exhaust vent
holes or anti-asphyxia valves needs to be prevented to avoid having an adverse effect
on the safety and quality of the therapy. Regularly inspect the vent holes and anti-
asphyxia valves to ensure they are kept clean, clear of blockages and are not
damaged.
• Only use compatible CPAP or bi-level therapy devices or accessories. The technical
specifications of the mask are provided for healthcare professionals to determine
compatible devices. Use in combination with incompatible medical devices can
decrease the safety or alter the performance of the mask.
• The mask is not suitable for patients requiring life support ventilation.
• Regularly clean your mask and its components to maintain the quality of your mask
and to prevent the growth of germs that can adversely affect your health.