Messages displayed by the BD Veritor System Reader are described in the table
below. For assay-specific messages, see the appropriate package insert.
Message
ANALYSIS
WAIT XXs
BARCODE
INVALID
RESULT
INVALID
CONTROL
INVALID
ERROR 5
BASELINE
INSERT
TEST
NO TESTS
REMAIN
READING
WAIT XXs
REMOVE
TEST
QC FAIL
QC PASS
READER
EXPIRED
Type
Status
Read in progress, where
XX is the total time
remaining
Error
The device barcode
cannot be read correctly
Error
Result invalid
Error
One or more of the
assay control lines was
outside of the normal
range
Error
The assay background
area was outside of the
normal range
Status
Ready to read
Error
The maximum number
of reader tests has been
performed
Status
Read in progress, where
XX is the total time
remaining
Status
A device is inserted in
the reader
Error
Result of a verification
test
Status
Result of a verification
test
Error
The operational life of
the reader is expired
Meaning
8
Action Needed
None
Repeat the assay. On
subsequent occurrences,
contact BD.
Repeat assay
On first occurrence, repeat
the assay. On subsequent
occurrences, contact BD.
Remove the device and
cycle the instrument power.
• If the instrument displays
an error during power-on,
contact BD.
• If no error message
appears, obtain an
additional sample and
repeat the assay. If the
error recurs, contact BD.
Insert a device
Contact your distributor to
order a new reader
None
Remove device
Contact BD
None
Contact your distributor to
order a new reader