Fail in power on or off
ON/OFF button stuck or
bad contact
Repeat displaying the
Query button stuck or
voltage
get short circuit
8. CALIBRATION
Please don't attempt to calibration without standard weight.
1) Turn on crane scale after display „0.000" press [ON/OFF] and [QUERY] until display
„------" loosen your hand it will display „SYS".
2) Press [QUERY] it will display "CAL-1".
3) Press [ON/OFF] will display capacity, you can input the poise weight by pushing but-
ton [QUERY] and [ON/OFF] as follows: push button [QUERY] to increase the flashing
number by one (circulate from 0 to 9) and push button [ON/OFF] to move the flashing
number toward left circular (e.g. if the poise weight is 3000kg, please input 03000).
4) After poise setting, you can press [on/off] for three seconds and it will display"C1",
suspend the poise, after the light of steady lamp is on, press [QUERY], the numbers
EN
is randomized. After the numbers steady it will display the weight of poise. Now the
calibration is finished
IF YOU NEED SETTING OF SYSTEM, PLEASE AFTER SETTING OF
SYSTEM THEN DO CALIBRATION.
check and replace the
button
check and replace the
button
36
SETTING OF SYSTEM
Turn on crane scale after display „0.000" press [ON/OFF] and [QUERY] until display
„------" loosen your hand it will display „SYS".
Press [QUERY] when it display „SCALE" press [ON/OFF] will display „00000" now has
enter into setting of capacity. Push button [QUERY] to increase the flashing number by
one (circulate from 0 to 9) and push button [ON/OFF] to move the flashing number
toward left circular (e.g. if capacity is 3000kg, you input 03000), then press [ON/OFF]
three seconds will display „SCALE", it means the capacity is confirm.
Press [QUERY] it will display „DEC", press [ON/OFF] will display „00000", Press
[QUERY] to move the radix point toward left. Then press [ON/OFF] to choose, it will
display „DEC"(e.g. if division is 0.5kg,you could choose 0000.0;if division is 1kg,you
could choose 00000).
Press [QUERY] it will display „PL-DE", press [ON/OFF] will display „N1",Press [QUE-
RY] TO CHOOSE"N1"or „N2"(N1 means the division is the unit from 1 to 9, N2
means is decade from 10 to 19),then press [ON/OFF] to exit, it will display „PL - DE".
If you only want one or two data, you want quit on halfway, press [ON/OFF] until
display "End" will back to weighing mode. Press [QUERY] it will display „CLASS", press
[ON/OFF] will display „1",Press [QUERY] to CHOOSE „1" „2" „5",then press [ON/
OFF] to exit, it will display „CLASS". If you only want one or two data, you want quit
on halfway, press [ON/OFF] until display "End" will back to weighing mode.
If you don't need calibration now you could turn off scale, then turn on scale to use.
If you need calibration Press [QUERY] it will display „CAL". Next step same as up
calibration.