Sample Identification - Xylem TitroLine 7000 Manual De Instrucciones

Ocultar thumbs Ver también para TitroLine 7000:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

196
The following filling options can be set for the dosing mode (Fig. 155):
Fig. 155
«off»
«always»
«intelligent before»
«intelligent after»

Sample identification

4.6.9
In the manual titration and in the preparation of solutions it is possible to input a sample identification (Fig. 156).
The possible input includes manual, automatic or no sample description at all.
Fig. 156
For a sample description of the «manual», a prompt for the sample description will always be displayed at the
start of the method (See also  3.6 Main Menu).
For an «automatic» sample description there will be selected a master description (e.g. Fig. 157 in the current
case this is water), which will then automatically be numbered starting on 01.
Fig. 157
After a new power-up, numbering will resume with 01.
filling it will not occur automatically after each dosing step.
filling will occur automatically after each dosing step.
a verification will be performed each time prior to the next dosing step in order to
determine whether the dosing step can still be made without a filling operation.
Should this prove to be impossible, the first thing to occur is filling, followed by the
dosing step.
a verification will be performed after the next dosing step to find out whether the next
dosing step can still be made without filling.