The following filling options can be set for the dosing mode (Fig. 125):
Fig. 125
«off»
«always»
«intelligent before»
«intelligent after»
Sample identification
4.6.7
In the manual titration and in the preparation of solutions it is possible to input a sample identification (Fig. 126).
The possible input includes manual, automatic or no sample description at all.
Fig. 126
For a sample description of the «manual», a prompt for the sample description will always be displayed at the
start of the method (See also 3.6 Main Menu).
For an «automatic» sample description there will be selected a master description (e.g. Fig. 127 in the current
case this is water), which will then automatically be numbered starting on 01.
Fig. 127
After a new power-up, numbering will resume with 01.
filling it will not occur automatically after each dosing step.
filling will occur automatically after each dosing step.
a verification will be performed each time prior to the next dosing step in order to
determine whether the dosing step can still be made without a filling operation.
Should this prove to be impossible, the first thing to occur is filling, followed by the
dosing step.
a verification will be performed after the next dosing step to find out whether the next
dosing step can still be made without filling.
163