6.3.6. Insert endodontic file into the canal and connect the file clip to the file (to ensure precise mea-
surements the file size should be adjusted to the canal diameter). Alternatively, use the touch
probe to provide electrical contact with the endodontic file.
The file image inside the tooth will appear (accompanied by a double beep signal). No file ima-
ge and no sound signal indicate a faulty connection. Check your connections, clean your file
clip/touch probe contact, moisten the canal, if necessary, and start again.
No other adjustments are necessary before starting measurements.
„ 6.4. Cable connection test
Connection test feature is included in CanalPro Apex Locator in order to
check the cables:
• Connect the file clip contact to the lip clip.
• "Connection" symbol
play, indicating proper connection (Pic. 2).
• If the symbol do not appear, the measurement cable or the file clip should
be replaced.
„ 6.5 Apex localization
6.5.1.
Coronal and Medial Zone
Slowly introduce the file into the canal.
The progression of the file inside the canal shown inside the tooth image on the left part of the
display (Pic. 3 to Pic. 5).
Picture 3
should appear on the status bar of the dis-
Picture 4
Picture 2
Picture 5
23