Medtronic MiniLink REAL-Time Manual Del Usaurio página 25

Ocultar thumbs Ver también para MiniLink REAL-Time:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Immunity Test
IEC 60601 Test
Electrostatic dis-
±6 kV indirect
charge (ESD)
 IEC
61000-4-2
 ±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
 IEC
 ±1 kV for in-
61000-4-4
put/output lines
Surge
±1 kV lines(s) to
line(s)
 IEC
61000-4-5
 ±2 kV line(s)
to earth
Voltage dips,
<5% U
short interrup-
dip in U
tions and volt-
cycle
age variations
on power supply
 40% U
lines
dip in U
cycles
 IEC
61000-4-11
 70% U
dip in U
cycles
 <5% U
(>95% dip in U
for 5 seconds
Power frequen-
3 A/m
cy (50/60 Hz)
magnetic field
 IEC
61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Level
Level
±6 kV indirect
(30-60% relative
humidity)
 ±30 kV air
(<5% relative
humidity)
Not applicable
 Not applicable
Not applicable
 Not applicable
(>95%
Not applicable
T
) for 0.5
T
(60%
T
 Not applicable
) for 5
T
(30%
T
) for 25
 Not applicable
T
T
 Not applicable
)
T
3 A/m
-23-
Electromagnetic Environment -
Guidance
The MiniLink transmitter should not
be affected by electrostatic dis-
charge that might occur under nor-
mal conditions of use.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.