Warning:
Cracking, flaking, or damage of the housing are signs of deterioration
and the performance of the device may be compromised. This may
affect the ability to properly clean and disinfect the transmitter. If these
signs are noted, stop using the device and call the Medtronic Diabetes
24 Hour HelpLine. The device must be discarded according to local
regulations for battery disposal (non-incineration).
To clean the transmitter:
 1
Wash your hands thoroughly.
 2
Attach the tester to the transmitter.
 3
If optional occlusive dressing was used and there is adhesive residue on the
transmitter, follow the instructions in Removing adhesive residue, on page 18.
 4
Rinse the transmitter under cool tap water for at least one minute and until visibly
clean. Make sure all hard-to-reach areas are rinsed completely.
 5
Prepare a mild liquid soap solution using one teaspoon (5 milliliters) of Ivory
soap per one gallon (3.8 liters) of room temperature tap water.
-10-
®
liquid