Warning:
If you see any body fluid in the connector opening, you must discard
the transmitter. Because the transmitter contains a battery, do not
discard in a bio-waste container. Instead, continue to clean and
disinfect the transmitter, and then discard according to local
regulations for battery disposal (non-incineration).
3
Attach the tester to the transmitter.
4
If optional occlusive dressing was used and there is adhesive residue on the
transmitter, follow the instructions in Removing adhesive residue, on page 18.
5
Rinse the transmitter under cool tap water for at least one minute and until any visible
debris is gone.
6
Prepare an enzymatic solution using two (2) ounces ENZOL Enzymatic Detergent per
gallon of water. Make sure to prepare a fresh solution for each use.
connector opening
connector pins
housing