Medtronic MiniLink REAL-Time Manual Del Usaurio página 21

Ocultar thumbs Ver también para MiniLink REAL-Time:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Warning:
For multiple-patient use, always clean and disinfect the transmitter
after removing it from the patient and before attaching it to the
charger. If blood comes in contact with any surface areas of the
charger, the contaminated device must be discarded. The charger
contains a battery which may explode upon incineration.
To clean the charger:
1
Wash your hands thoroughly.
2
Use a damp cloth with mild cleaning solution, such as a dishwashing detergent,
to clean any dirt or foreign material from the outside of the charger. Never use
organic solvents, such as paint thinner or acetone, to clean the charger.
3
Place the charger on a clean, dry cloth and air dry for two to three minutes.
Troubleshooting
Question: Why do I see quick flashing red lights on the charger?
about 2 secs
Answer: The transmitter battery is very low. Leave the transmitter on the charger for
eight hours to completely recharge. If the red light is still flashing after eight hours,
leave your transmitter on the charger for 24 hours. If the red light is still flashing after
it has been charging continuously for 24 hours, contact the 24 Hour HelpLine. It may
be time to replace your transmitter.
Question: Why did the flashing green charger light turn off and a longer flashing red
charger light turn on during charging?
about 2 secs
Answer: Your charger battery is low. Make sure the transmitter is not connected to
the charger and replace the charger battery with a new AAA battery.
about 2 secs
about 2 secs
-19-
about 2 secs
about 2 secs
Tabla de contenido
loading

Tabla de contenido