SECTION 1—GENERAL GUIDELINES
For optimum performance, Invacare recommends that each
concentrator be on and running for a minimum of 5 minutes.
Shorter periods of operation may reduce maximum product
life.
The XPO
is not designed to be used with a humidifier. Use of
2
this device with a humidifier may impair performance and/or
damage the equipment.
Use of any means or device to reduce fire propagation in the
supply accessories is contraindicated and may cause
improper oxygen therapy delivery.
When your automobile is turned Off, disconnect the car
accessory power supply and remove the XPO
automobile. NEVER allow the XPO
or cold automobile or in other similar, high or low,
temperature environments. Refer to Typical Product
Parameters on page 14.
DO NOT operate in temperatures below 41° F (5°C) or above
104° F (40° C) for extended periods of time.
The XPO
MUST be used in an upright position.
2
Invacare recommends the XPO
Maintenance
The XPO
was specifically designed to minimize routine
2
preventive maintenance. Only professionals of the healthcare
field or persons fully conversant with this process such as
factory trained personnel should perform preventive
maintenance or performance adjustments on the oxygen
concentrator.
Invacare® XPO
™
2
to be stored in a very hot
2
not be used in the rain.
2
8
from the
2
Part No 1160887