Invacare XPO2 Manual De Usuario página 15

Ocultar thumbs Ver también para Invacare XPO2:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 100
Altitude:
Oxygen Concentration:*
*Based on an atmospheric
pressure of 14.7 psi (101 kPa) at
70°F (21°C)
Conserver Trigger Sensitivity:
Factory set - no adjustment
Pressure activated
Conserver Trigger Delay:
Factory set - no adjustment
Conserver Bolus Delivery:
Fixed minute volume
Conserver Breath Rate Capacity:
Dimensions:
Weight:
Battery Duration (times are
approximate):
Battery Recharge Time:
NOTE: Recharge time increases if
battery is charging while unit is
running.
Part No 1160887
SECTION 5—TYPICAL PRODUCT PARAMETERS
Up to 10,000 ft (3046 m) above sea level
87% to 95.6%, after initial warm-up
period (approximately 5 minutes)
< 0.20 cmH
O pressure drop
2
All settings - using 7 ft (2.1 m) cannula
50 mSec
Nominal value - using 7 ft (2.1 m) cannula
Setting 1: 320 cc
Setting 2: 480 cc
Setting 3: 670 cc
Setting 4: 800 cc
Setting 5: 900 cc
Total cc's ± 75 cc delivered per minute
Up to, and including, 35 BPM without
reduction of bolus minute volume
10 in high x 7 in wide x 4 in deep (25.4
cm high x 17.8 cm wide x 10.2 cm wide)
6.0 lbs (2.7 kg); 1.3 lbs (0.68 kg) for
supplemental battery
Setting 1 = 3.5 hrs
Setting 2 = 2.5 hrs
Setting 3 = 2.0 hrs
Setting 4 = 1.5 hrs
Setting 5 = 1.0 hrs
4 hours