Step 3) Adjusting The Torque - Nouvag MD 20 Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

MD20
selectable rotational
speed range
2)
Select desired rotational speed with „-" or „+".
3)
No longer activate "Speed" keys „-" and „+" → after approx. 2s, the display jumps back into normal
representation.
→ In normal representation, the selected mode now appears on the top right.
With set fixed keys, the rotational speed is limited to exactly one parameter.
6.5.3.

Step 3) Adjusting the torque

After selecting the rotational speed, the torque can be selected from the corresponding torque range.
Depending on the rotational speed, the torque limitations AL and AS can also be used.
For information on AL and AS mode, see chapter „6.6 AL torque limitation
(Automatic Limiter)" and chapter „6.7 AS torque limitation (Automatic Stopper)"
Adjust torque:
1)
Press "Torque" key „-" or „+" → Torque selection menu appears:
selectable torque
range
selected
torque
2)
Select desired mode with „-" or „+".
3)
No longer activate "Torque" key „-" and „+" → after approx. 2s the display jumps into normal
representation.
→ In normal representation, the selected mode now appears on the bottom left.
With set fixed keys, the torque is limited to exactly one parameter.
10
selected
rotational
speed
NOUVAG AG
No. 31842 14/06
Tabla de contenido
loading

Este manual también es adecuado para:

31842

Tabla de contenido