About this Publication
Activate a New Detector
This publication is an operator guide for BW Clip
Real Time portable single-gas detectors
1. Move to a normal atmosphere (20.9% v/v O
manufactured by BW Technologies by
2. Press and hold the pushbutton until a 5 second countdown is displayed, and then
Honeywell. It is intended for use with the
following models:
continue to hold until the countdown is complete.
3. When the countdown is complete, the LCD and LEDs turn on and then turn off. The
• 24- and 36-month H 2 S and CO Detectors
detector performs an internal diagnostic test.
24-month O
detector
2
4. When the diagnostic test is successful, the alarm setpoints are displayed and the LEDs
• 24-month SO
detector
2
Copyright, Notice, Trademarks
flash.
5. The sensor stabilization countdown is displayed. During the countdown, the detector
While this information is presented in good faith
and believed to be accurate, BW Technologies
vibrates continuously for 20 seconds. The time required to stabilize varies depending
by Honeywell (BW) disclaims the implied
on sensor type. When the countdown reaches 0, the LEDs flash and the audible alarm
warranties of merchantability and fitness for a
beeps.
particular purpose and makes no express
6. The detector is in normal operating mode when the gas type and concentration are
warranties except as may be stated in its written
displayed.
agreement with and for its customers.
In no event is BW liable to anyone for any
Normal Operating Mode
indirect, special or consequential damages. The
content in this document is subject to change.
When the detector is in normal operating mode, the type of gas
Contact Us
detected is permanently displayed. The detected concentration of the
Corporate Headquarters
gas is displayed until it is disrupted by a pushbutton action, gas alarm,
BW Technologies by Honeywell
or error event.
Suite 110 4411-6 Street SE
Calgary, Alberta
Operating Life
Canada, T2G 4E8
Toll-free: 1-800-663-4164
When the detector is activated for the first time, the operating life
Americas
countdown is displayed. Press the pushbutton to display remaining
BW Technologies by Honeywell
operating life.
405 Barclay Blvd.
Lincolnshire, IL
USA 60069
90 Days or Less Remaining
Toll-free: 1-888-749-8878
When the operating life countdown reaches 90 days or less before
Europe
expiry, the countdown display changes to days. The countdown
BW Technologies by Honeywell
Javastrasse 2
displays the remaining operating life in days until the countdown
8604 Hegnau
reaches 24 hours or less before expiry.
Switzerland
24 Hours or Less Remaining
Toll-free: 00800-333-222-44
When the operating life countdown reaches 24 hours or less before
Other
BW Technologies by Honeywell
expiry, the countdown display changes to hours, and the EXPIRY!
Toll-free: 1-403-248-9226
warning and non-compliance symbol are displayed.
www.honeywellanalytics.com
End of Operating Life
Warranty Registration
When the operating life countdown ends, the detector deactivates and
www.honeywellanalytics.com/support/
detector safety functions are disabled. It is possible to retrieve event
product-registration
logs for a limited time after expiry. The EXPIRY! warning and
Maximum Operating Life
non-compliance symbol are displayed for up to 30 days after expiry.
24-
month Detector
2 years after activation, assuming 2 minutes
Internal Diagnostic Tests
of alarm time per day.
Note: 24-month H 2 S and CO detectors that
Activated detectors automatically perform one internal diagnostic test every 24 hours. If the
use hibernation may extend the service
internal diagnostic test fails, the diagnostic fail-safe begins.
period of the detector up to an additional
Diagnostic Fail-Safe
year. The service life will end once the
detector reaches 24 months of actual
1. The EXPIRY! warning and non-compliance symbol are displayed.
operation. Hibernation provides a maximum
service life of 36 months.
2. The detector beeps and vibrates.
3. The LEDs flash 2 times per second for 15 seconds.
36-
month Detector
4. OFF or an error code is displayed. The detector is deactivated.
3 years after activation, assuming 1 minute
of alarm time per day.
It is possible to retrieve the event logs for a limited time after expiry. If you are unable to
Detection Range
retrieve the event logs, contact BW.
H
S
0 to 100 ppm
2
Alarms and Warnings
CO
0 to 300 ppm
2
O
0 to 25% v/v O
An alarm is initiated when the sensor is exposed to a gas concentration that exceeds alarm
2
2
SO
0 to 100 ppm
2
setpoints. The alarm persists until the alarm gas concentration returns to an acceptable
Factory Alarm Setpoints
range. Battery life decreases rapidly when the detector is in alarm condition.
Low Alarm
High Alarm
Low Alarm
H
S
10 ppm
15 ppm
2
Audible: 1 beep per second
CO
35 ppm
200 ppm
Visual: 1 flash per second
O
19.5% v/v O
23.5% v/v O
2
2
SO
5 ppm
10 ppm
Vibration: 1 vibration per second
2
Use Fleet Manager II via an IntelliDoX station to
adjust alarm setpoints. For more information,
High Alarm
refer to the operator manuals for Fleet
Audible: 2 beeps per second
Manager II software version 4.2 and the
IntelliDoX automatic test and calibration
Visual: 2 flashes per second
station.
Ratings and Certifications
Vibration: 2 vibrations per second
The BW Clip Real Time is in conformity with
the following standards:
UL 913 8th Edition
UL 60079-0:2013,
UL 60079-11:2013
SINGLE-GAS DETECTOR
SINGLE-GAS DETECTOR
2
2
CSA C22.2 No. 157-92:2012,
CSA C22.2 No. 60079-0:2011,
CSA C22.2 No. 60079-11:2014
) that is free of hazardous gas.
EN 60079-0:2012 +A11:2013,
2
EN 60079-11:2012
IEC 60079-0:2011
IEC 60079-11:2011
ABNT NBR IEC 60079-0:2008
ABNT NBR IEC 60079-11:2009
Classified by UL to both US and
Canadian Standards as intrinsically safe
UL
for Class I, Division 1, Group A, B, C, D
and Class I, Zone 0, Group IIC.
ATEX
CE 0539 II 1G
Ex ia IIC T4 Ga IP66/67
DEMKO 14 ATEX 1356
CE
European Conformity
EU Declaration of Conformity
www.honeywellanalytics.com
IECEx
Ex ia IIC T4 Ga IP66/67
IECEx UL 14.0063
INMETRO
Ex ia IIC T4 Ga
DNV 15.0024
KTL
1
15-KA4BO-0306
Specifications
Shelf Life
H
S
One (1) year before activation
2
CO
One (1) year before activation
O
Six (6) months before activation
2
SO
One (1) year before activation
2
Instrument Weight alligator clip included
92 grams (3.2 ounces)
Instrument Dimensions alligator clip
included
41 x 50 x 87 millimeters
(1.6 x 2.0 x 3.4 inches)
Operating Temperatures
H
S
-40° to +50°C (-40° to +122°F)
2
CO
-30° to +50°C (-22° to +122°F)
O
-20° to +50°C (-4° to +122°F)
2
SO
-30° to +50°C (-22° to +122°F)
2
Internal Vibrating
Operates to -10°C (+14°F)
Operating Humidity
5% to 95% relative humidity
(non-condensing)
Audible Alarm
≈ 95 dB at 10 cm (3.9 in)
Visual Alarm
Flashing, wide-angled alarm lens with red
LEDs plus alarm LCD readout
Display
Liquid crystal display (LCD)
Sensor Type
Electrochemical cells
Battery
Lithium, non-replaceable
Ingress Protection
2
IP 66/67
Alarm Setpoints
Instant low and instant high.
Calibration
To ensure accurate calibration, BW
Technologies recommends using a premium-
grade calibration gas approved by the National
Institute of Standards and Technology (NIST).
Do not use a gas cylinder beyond its expiry date.
BW Technologies recommends you calibrate the
detector at least once every 180 days depending
http://www.honeywellanalytics.com/~/media/
1
honeywell-analytics/products/bw-clip/docu-ments/certificates/
korea-certi ication_bw-clip-15ka-4bo0306
on use and sensor exposure to poisons and
contaminants. Calibrate only in a safe area that
Over-Limit and Under-Limit Alarms
is free of hazardous gas.
Allowable Calibration Gas Mixtures
If the gas concentration is beyond the sensor range, OL (over-limit) or
H
S
5 to 50 ppm
2
-OL (under-limit) is displayed.
CO
25 to 250 ppm
2
SO
5 to 70 ppm
Audible: 2 beeps per second
2
O
5.0 % to 19.0 % v/v O
2
2
Visual: 2 flashes per second
For O
detectors, perform the Zero procedure
2
once every 24 hours or when the Automatic
Vibration: 2 vibrations per second
Zero Reminder is displayed.
Intended Use
This product is classified for use in hazardous
Automatic Zero Reminder for O
atmospheres that are not more than
21% v/v O
For O
models only, the Automatic Zero Reminder is displayed when
2
2
more than 24 hours have elapsed since the last successful zero
Recycling
procedure was performed.
This instrument contains a lithium battery. Do not
mix with the solid waste stream. Spent batteries
should be disposed of by a qualified recycler or
Peak Readings
hazardous materials handler.
The peak readings symbol is displayed when the sensor is exposed to
a CAUTION
a gas concentration that exceeds the alarm setpoints. It is no longer
Products may contain materials that are
displayed when more than 24 hours have passed since the last alarm,
regulated for transportation under domestic and
or when a successful bump test is performed via an IntelliDoX or
international dangerous goods regulations.
MicroDock II station.
Return product in compliance with appropriate
dangerous goods regulations. Contact freight
Non-compliance Symbol and LED Indicators
carrier for further instructions.
Limited Warranty and
The non-compliance warning symbol is displayed when:
Limitation of Liability
an internal diagnostic test fails;
BW Technologies (BW) warrants this product to
a bump test, calibration or zero procedure fails;
be free from defects in material and workmanship
a bump test or calibration is due, if enabled;
under normal use and service for a period of two
the sensor is exposed to a gas concentration that exceeds alarm setpoints, if enabled.
or three years (depending upon detector),
beginning on the date of activation. 24-month
The non-compliance warning symbol is no longer displayed when more than 24 hours have
H
S and CO detectors are covered for up to an
2
passed since the last alarm, or when a successful bump test or calibration is performed
additional 12 months when hibernation is used,
limited by a total of 24 months of detector
(manual calibration or via an IntelliDoX or MicroDock II station). The non-compliance
operation.
warning symbol is permanently displayed when the service-life countdown reaches 24 hours
This Warranty is valid only if the detector is
or less before expiry, and for up to 30 days after expiry.
activated by the date on the package. This
warranty extends only to the sale of new and
When Non-compliance after gas exposure is enabled through Fleet Manager II via
unused products to the original buyer. BW's
an IntelliDoX station, the LEDs flash in an alternating pattern when the non-compliance
warranty obligation is limited, at BW's option, to
refund of the purchase price, repair, or
warning symbol is displayed. When the symbol is first displayed, the top LEDs flash. After
replacement of a defective product that is
30 seconds, the side LEDs flash. After 30 seconds, the pattern repeats until the non-
returned to a BW authorized service center within
compliance warning symbol is no longer displayed.
the warranty period.
NOTE: When the detector is bump tested via a MicroDock II station, an abnormal test
result may occur if non-compliance LEDs are flashing. To suspend the flashing for up to 10
minutes, press and release the pushbutton. When the flashing stops, insert the detector and
perform the test. If the bump test fails, the LEDs resume flashing.
Recall Events and Settings
When the detector is in normal operating mode, press the pushbutton at any time to scroll
through events and settings that are not empty or disabled through Fleet Manager II via an
IntelliDoX station. BW Clip Real Time flashes, vibrates and beeps.
While an event or setting is displayed, press the pushbutton to scroll to the next available
event or setting. If you do not press the pushbutton, or if you have reached the last available
event or setting, the detector returns to normal operation. Events and settings are displayed
in this order:
1. Peak Reading, if available, and when a maximum gas exposure event has occurred
within the past 24 hours.
2. Time of Peak Reading, if Real-time Clock display is enabled.
3. Operating Life, the remaining operating life of the detector in months, days or hours.
4. Low Alarm Setpoint
5. High Alarm Setpoint
6. Real-time Clock, if enabled.
7. Firmware Version
8. Next Bump Due, if enabled.
9. Next Calibration Due, if enabled.
10. Hibernation, for 2-year H
S and CO models only.
2
In no event shall BW's liability hereunder exceed
the purchase price actually paid by the buyer for
the Product.
This warranty does not include:
• fuses, disposable batteries or the routine
replacement of parts due to the normal
wear and tear of the product arising from
use;
any product which in BW's opinion,
has been misused, altered, neglected
or damaged by accident or abnormal
conditions of operation, handling or use; or
any damage or defects attributable to repair of
the product by any person
Models
other than an authorized dealer, or the
2
installation of unapproved parts on the product.
The obligations set forth in this warranty are
conditional on:
1) proper storage, installation, calibration, use,
maintenance and compliance with the
product manual instructions and any other
applicable recommendations of BW;
2) the buyer promptly notifying BW of any
defect and, if required, promptly making the
product available for correction.
No goods shall be returned to BW
until receipt by the buyer of shipping
instructions from BW; and
3) the right of BW to require that the buyer
provide proof of purchase such as the
original invoice, bill of sale or packing slip to
establish that the product is within the
warranty period.
THE BUYER AGREES THAT THIS WARRANTY
IS THE BUYER'S SOLE AND EXCLUSIVE
REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS
OR IMPLIED, INCLUDING BUT NOT LIMITED
TO ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. BW SHALL NOT BE
LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES
OR LOSSES, INCLUDING LOSS OF DATA,
WHETHER ARISING FROM BREACH OF
WARRANTY OR BASED ON CONTRACT, TORT
OR RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow
limitation of the term of an implied warranty, or
exclusion or limitation of incidental or
consequential damages, the limitations and
exclusions of this warranty may not apply to
every buyer. If any provision of this warranty is
held invalid or unenforceable by a court of
competent jurisdiction, such holding will not affect
the validity or enforceability of any other
provisions.
0919 IH-6936VND
Tabla de contenido
loading

Tabla de contenido