BW Clip Real Time
2 OR 3 YEAR H
S, CO, O
, SO
2
2
BW Clip Real Time is a portable, single-gas detector manufactured by BW Technologies by
Honeywell (BW). BW Clip Real Time continuously displays the concentration of a specific
gas in the ambient environment, and activates an alarm when concentrations exceed alarm
setpoints. It is your responsibility to respond appropriately to the alarms.
BW Clip Real Time is a member of the GasAlert family of products. It is compatible with
IntelliDoX and MicroDock II automatic test and calibration stations, as well as Fleet Manager
II software version 4.2 or higher. For more information, visit the product website at www.
honeywellanalytics.com
Safety Information: Read First
1. Substitution of components may impair Intrinsic Safety.
2. BW recommends performing a bump test prior to each day's use to confirm sensor
response and alarm activation by exposing the detector to a concentration of target
gas that exceeds the low alarm set point. BW also recommends performing a bump
test if the detector has been subjected to physical impact, liquid immersion, an Over
Limit alarm event, custody changes, or anytime the detectors performance is in doubt.
Calibrate the detector after a failed bump test. Always bump test and calibrate in a fresh
air environment.
3. Detectors in hibernation mode are not functional.
1. Activate the detector before the activation date on the package.
2. This product is a gas detector, not a measurement device.
3. Ensure that the sensor grill is free of dirt, debris, and is not obstructed.
4. Clean the exterior with a soft, damp cloth.
5. For optimal performance, periodically zero the sensor in a normal atmosphere (20.9%
v/v O
) that is free of hazardous gas.
2
6. Portable safety gas detectors are life safety devices. Accuracy of ambient gas
reading(s) is dependent upon factors such as accuracy of the calibration gas standard
used for calibration and frequency of calibration. BW recommends performing a
calibration at least once every 180 days (6 months).
Informations de sécurité – À lire au préalable
AVERTISSEMENT
1. Le remplacement d'un composant de l'appareil peut compromettre la sécurité
intrinsèque du détecteur.
2. Avant chaque utilisation quotidienne, BW recommande d'effectuer un test fonctionnel
afin de vérifier la réaction du capteur et l'activation de l'alarme, en exposant le détecteur
à une concentration de gaz cible supérieure au seuil d'alarme basse. BW recommande
également d'effectuer un test fonctionnel si le détecteur a été soumis à un impact
physique, à une immersion dans du liquide, à un événement d'alarme Dépassement
de la limite, à des changements d'utilisateur ou chaque fois que les performances
du détecteur sont mises en doute. Étalonnez le détecteur après l'échec d'un test
fonctionnel. Effectuez toujours un test fonctionnel et étalonnez dans un environnement
avec de l'air frais.
3. Détecteurs en mode veille prolongée ne sont pas fonctionnels.
MISE EN GARDE
1. Activez le détecteur avant la date d'activation sur l'emballage.
2. Ce produit est un détecteur de gaz, pas un dispositif de mesure.
3. Veillez à assurer que la grille du capteur est exempt de saleté, les débris, et ne soit pas
obstrué.
4. Nettoyez l'extérieur de l'appareil à l'aide d'un chiffon doux et humide.
5. Pour des performances optimales, zéro périodiquement le capteur dans une
atmosphère normale (20,9% v/v O
2
6. Les détecteurs portables de gaz de sécurité sont des dispositifs de sécurité des
personnes. La précision des valeurs de gaz ambiant dépend de divers facteurs,
tels que la précision de la norme de gaz d'étalonnage utilisée pour l'étalonnage et la
fréquence d'étalonnage. BW recommande d'effectuer un étalonnage au moins une fois
tous les 180 jours (6 mois).
SINGLE-GAS DETECTOR
2
) qui est exempte de gaz dangereux.
Part Number 50115903-101 | OM-EN-BWCRT-A6 | BW Clip Real Time Operator Manual
Getting Started
LED
LED
visual alarms
visual alarms
LCD
LCD
Audible
Audible
alarm
alarm
Bubble label
Bubble label
with product name
with product name
and gas type
and gas type
LCD and Symbols
Pushbutton prompt. When this symbol is displayed, press and hold
the pushbutton until the symbol disappears.
BW Clip Real Time O
measured as percent by volume.
BW Clip Real Time H
concentration is measured in parts per million.
Low alarm and high alarm. These symbols are displayed when a
gas concentration that exceeds alarm setpoints is detected.
Non-compliance warning. This symbol is displayed when a
diagnostic or compliance test fails; when the operating life
countdown reaches 24 hours or less; and for up to 30 days after
operating life expiry. It is also displayed when certain notifications
are enabled through Fleet Manager II via an IntelliDoX station.
Real-time clock. This symbol is displayed when real-time clock is
enabled through Fleet Manager II via an IntelliDoX station.
End of operating life warning. This symbol is displayed when the
operating life countdown reaches 24 hours or less.
Peak reading. This symbol is displayed when a maximum gas
exposure event has occurred within the past 24 hours.
Bump test reminder. If the detector is configured to display a bump
test reminder, this symbol is displayed when a bump test is due.
indicates a failed bump test.
This symbol is displayed when more than 24 hours have elapsed
since the last successful zero procedure was performed (O
models) or indicates a sensor calibration is due (H
models).
indicates a failed calibration (H
failed zero procedure (O
Operating life remaining in months (MM) or days (DD).
Infrared port
Pushbutton
Pushbutton
Sensor and
Sensor and
sensor grill
sensor grill
Product label
models only. Oxygen concentration is
2
S, CO and SO
models only. Toxic gas
2
2
S, CO and SO
2
S, CO and SO
models) or a
2
2
models).
2
0919 IH-6936VND
Alligator
clip
2
2
Tabla de contenido
loading

Resumen de contenidos para Honeywell BW CLIP REAL TIME

  • Página 1 It is your responsibility to respond appropriately to the alarms. BW Clip Real Time is a member of the GasAlert family of products. It is compatible with IntelliDoX and MicroDock II automatic test and calibration stations, as well as Fleet Manager Alligator II software version 4.2 or higher.
  • Página 2 36 months. 2. The detector beeps and vibrates. 5% to 95% relative humidity IntelliDoX station. BW Clip Real Time flashes, vibrates and beeps. 3. The LEDs flash 2 times per second for 15 seconds. month Detector (non-condensing) While an event or setting is displayed, press the pushbutton to scroll to the next available 4.
  • Página 3: Optional Settings

    24 hours or less before expiry. passed since the last alarm, or when a successful bump test or calibration is performed By default, BW Clip Real Time is configured to use the following calibration gas mixtures: 24 Hours or Less Remaining Time may be displayed in 12-hour or 24-hour clock format.
  • Página 4 Ratings and Certifications Do not use a gas cylinder beyond its expiry date. BW Technologies recommends you calibrate the The BW Clip Real Time is in conformity with detector at least once every 180 days depending the following standards: UL 913 8th Edition UL 60079-0:2013, http://www.honeywellanalytics.com/~/media/...
  • Página 5: Información Sobre Seguridad: Leer Primero

    Número de pieza 50115903-067 | OM-ES-PALO-A8 | Manual para el operador de BW Clip Puesta en marcha BW Clip es un detector portátil de un solo gas fabricado por BW Technologies de Honeywell (BW). BW Clip mide continuamente la concentración de un gas Alarmas visuales Alarmas visuales específico en el entorno ambiental y activa una alarma cuando l s concentraciones...
  • Página 6 BW Clip es un detector portátil de un solo gas fabricado por BW Technologies de Honeywell (BW). BW Clip mide continuamente la concentración de un gas 1. Ubíquese en un lugar donde haya una atmósfera normal (20,9 % v/v de O Se inicia una alarma cuando se expone el sensor a una concentración de gas...
  • Página 7: Registros De Eventos

    Puesta en marcha BW Clip es un detector portátil de un solo gas fabricado por BW Technologies de Honeywell (BW). BW Clip mide continuamente la concentración de un gas Cuando el detector se encuentre en modo de operación normal, presione el botón Utilice Fleet Manager II por medio de una estación IntelliDoX para habilitar los...
  • Página 8: Especificaciones

    5 % a 95 % de humedad relativa (sin buena fe y se presupone su exactitud, Clasificaciones y Esta garantía no incluye: condensación) BW Technologies de Honeywell (BW) renuncia certificaciones • fusibles, baterías desechables o el Alarma sonora a las garantías implícitas de comerciabilidad e...
  • Página 9: Mise En Route

    Référence 50115903-067 | OM-FR-PALO-A8 | Manuel de l'opérateur de BW Clip Mise en route BW Clip est un détecteur monogaz portable fabriqué par BW Technologies by Honeywell (BW). Le détecteur BW Clip mesure en continu la concentration d'un Alarmes visuelles Alarmes visuelles à...
  • Página 10: Mode De Fonctionnement Normal

    Activer un nouveau détecteur BW Clip est un détecteur monogaz portable fabriqué par BW Technologies by Alarmes et avertissements Honeywell (BW). Le détecteur BW Clip mesure en continu la concentration d'un Alarmes visuelles Alarmes visuelles 1. Déplacez le détecteur dans une atmosphère normale (20,9 % v/v O ) ne Une alarme est déclenchée lorsque le capteur est exposé...
  • Página 11 Rappeler des événements et des paramètres BW Clip est un détecteur monogaz portable fabriqué par BW Technologies by Paramètres facultatifs Honeywell (BW). Le détecteur BW Clip mesure en continu la concentration d'un Alarmes visuelles Alarmes visuelles Lorsque le détecteur est en mode de fonctionnement normal, appuyez sur Utilisez le logiciel Fleet Manager II via un système IntelliDoX pour activer les...
  • Página 12 15 ppm à l'acheteur d'origine. Les obligations de BW au -20 à +50 °C (-4 à +122 °F) par BW Technologies by Honeywell. Il est destiné 35 ppm 200 ppm titre de la garantie se limitent, au choix de BW, à...

Tabla de contenido