Optika B-510 Serie Manual De Instrucciones página 63

Ocultar thumbs Ver también para B-510 Serie:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

5.
Using the focus screw of the collector lens
①, put the image of the arc projected onto
the paper. The light spot must be brighter
and sharper as possible. (Fig. 44)
Using the centering screws ② on the side
6.
of the lamp housing, center the image of
the arc. (Fig.44-45)
7.
Using the focusing screw of the collector
lens ① enlarge the image until a
homogeneous illumination is achieved.
(Fig.46). At this point, insert an objective
into the optical path and, looking into
the eyepieces, optimize the illumination
always using the screws ① and ②.
8.
After replacing the exhausted bulb, reset
the time counter on the power supply by
pressing the "Reset" button ①. (Fig.47)
Page 63
Fig.44
Fig.45
Fig.46
Fig.47
Tabla de contenido
loading

Tabla de contenido