► B-510FL ONLY
•
Disconnect all electrical cables before installing or replacing the bulb.
•
The bulb has an anode and a cathode of different sizes. Respect the polarity during assembly, respecting
the bulb dimensions.
•
Do not touch the bulb of the lamp with bare hands to leave no traces of grease on the bulb. If this happens,
clean the bulb with a soft cloth before turning on the lamp.
•
The lamp has an average life of about 200-250 hours: a time counter and a voltage indicator are shown on
the lamp power supply. Replace the lamp when the hour count exceeds 250 or if the voltage drops below
4.5A.
•
During use, the lamp, the lamp housing and the surrounding environment become hot.
•
Before replacing the lamp, switch off the power supply, disconnect all cables and wait for the lamp and the
lamp housing to cool.
•
After switching on the lamp, wait at least 10-15 minutes before switching it off.
•
After switching off the lamp, wait for 5-10 minutes before switching it on again so that the mercury vapors
have time to condense.
•
The lamp contains ultraviolet radiation that could be harmful to eyes and skin. Always look at the lamp arc
through the provided orange screen.
►
B-510FL ONLY
1. Open the lamp housing using the door
lock screw ① and remove the lamp holder.
(Fig.34)
Remove the plastic block ② from the
2.
lamp holder (or the exhausted lamp in
case of replacement) by loosening the
two locking screws ③. (Fig.35)
IInsert the mercury bulb ④ (respect the
3.
polarity of the bulb), tighten the locking
screws and refit the lamp holder inside
the lamp housing. (Fig. 36)
Page 60
①
Fig.34
③
②
③
Fig.35
④
Fig.36