Disconnect all electrical cables before installing or replacing the bulb.
The bulb has an anode and a cathode of different sizes. Respect the polarity during assembly, respecting
the bulb dimensions.
Do not touch the bulb of the lamp with bare hands to leave no traces of grease on the bulb. If this happens,
clean the bulb with a soft cloth before turning on the lamp.
The lamp has an average life of about 200-250 hours: a time counter and a voltage indicator are shown on
the lamp power supply. Replace the lamp when the hour count exceeds 250 or if the voltage drops below
4.5A.
During use, the lamp, the lamp housing and the surrounding environment become hot.
Before replacing the lamp, switch off the power supply, disconnect all cables and wait for the lamp and the
lamp housing to cool.
After switching on the lamp, wait at least 10-15 minutes before switching it off.
After switching off the lamp, wait for 5-10 minutes before switching it on again so that the mercury vapors
have time to condense.
The lamp contains ultraviolet radiation that could be harmful to eyes and skin. Always look at the lamp arc
through the provided orange screen.
B-510FL ONLY
1. Open the lamp housing using the door
lock screw ① and remove the lamp holder.
(Fig.34)
Remove the plastic block ② from the
2.
lamp holder (or the exhausted lamp in
case of replacement) by loosening the
two locking screws ③. (Fig.35)
IInsert the mercury bulb ④ (respect the
3.
polarity of the bulb), tighten the locking
screws and refit the lamp holder inside
the lamp housing. (Fig. 36)