Preventive Maintenance; Accessories And Consumables - Huntleigh sonicaidONE Guia Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 68
SonicaidOne User Guide
13
Return of goods
To return a SonicaidOne for repair:
1 Clean the product, and return in suitable packaging.
2 Attach a Certificate of Decontamination or other written declaration that the
product has been cleaned to the outside of the package.
3 Mark the package 'Service Department, SonicaidOne'.

Preventive maintenance

Checking for damage
Once a month, check the SonicaidOne probe for damage or cracks which may
allow the ingress of liquids or gel.
Cleaning and disinfecting the probe
After each use, clean the probe with a cloth which has been dampened in soap
or detergent solution. Dry it with a clean cloth. Disinfect it using an alcohol-
impregnated wipe (70% ethanol or isopropyl alcohol).
Cleaning the main unit
In normal use the main unit needs occasional cleaning, and will not need
disinfection. Once a week, clean the case with a cloth which has been
dampened in soap or detergent solution, then dry it with a clean cloth.
If soiled, disinfect using an alcohol-impregnated wipe (70% ethanol or isopropyl
alcohol).
Note : the SonicaidOne is an electronic device and therefore must not
be sterilized.

Accessories and consumables

Accessories:
230V UK charger including rechargeable nickel-cadmium batteries
230V Euro charger including rechargeable nickel-cadmium batteries
110V US charger including rechargeable nickel-cadmium batteries
Headset
The only accessory which can be connected to the SonicaidOne is the optional
headset. No other equipment should be connected.
Tabla de contenido
loading

Tabla de contenido