QC lockout configuration

Touch
in the configuration menu to enter the QC lockout setting menu.
Touch
Touch
Touch
1

Select assay type

Touch the arrow in the window to open the drop down menu to select the assay type.
Touch the assay to select.
2

QC lockout

Select
Select
control level C I OR C II, to reset the QC lockout interval.
Select
both control level C I AND C II, to reset the QC lockout interval.
Touch
3

QC lockout interval

Select
Select
Touch
Touch
4

Control lot database

To add a control to the control lot database the Afinion Control Data is required.
The Afinion Control Data is a numeric data string which contains all lot specific data:
• Afinion control lot number
• Control type (assay)
• Control level (C I or C II)
The Afinion Control Data and its accompanying barcode is found in the Afinion Control kit
package insert. If the Afinion Control Data is not available, contact your local supplier.
Touch
(recommended), scan the barcode.
The Afinion Control Data may also be entered before, during or after a control run. The control lot will
automatically be stored in the database. See page 25.
Select lot number and touch
When a control lot has reached its expiry date, the control will automatically be deleted from the instrument
control database. The control lot database can store 100 control lots.
to configure QC lockout for the assay selected.
to configure QC lockout interval.
to view/add/delete stored control lots in the control lot database.
to disable the QC lockout function. No QC runs will be required for this assay
to enable the QC lockout function. It is required to run ONE passed control,
to enable the QC lockout function. It is required to run TWO passed control,
to confirm and return to the previous view.
to configure QC lockout interval by number of runs.
to configure QC lockout interval by hours.
to enter/edit number of runs/hours to QC lockout.
displays the number of runs/hours configured in the QC lockout interval.
to confirm and return to the previous view.
and either manually enter the Control Data or if a barcode reader is connected to the Analyzer
• Control expiry date
• Acceptable control range
• CRC (check sum to validate the previous data)
to delete a control from the list.
Configuration
EN | 17