Operation; Power On/Off The Oximeter - Gima OXY 110 Guía De Uso

Ocultar thumbs Ver también para OXY 110:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

47
• The infrared temperature probe will switch to standby state automatically if
there is no operation for 1 minute. If a further measurement is needed, press
the measuring key and repeat step 2 and step 3.
• Normal body temperature varies depending on the position/area the measure-
ment is taken from. The following table shows the varying temperature ranges
of the different body positions.
Temperature varying range at different body positions:
Arm
Oral
Rectal
Ear
Safety Instruction for Temperature Measurement
 Do NOT take a measurement when the patient is moving.
 Patients with tympanitis or otitis problems should NOT use this device.
 When the infrared temperature probe is connected to the device, the probe will
consequently be at power-on status, therefore pressing the power on/off key
on the temperature probe will not cause any effect.

4 Operation

4.1 Power on/off the Oximeter

• Long pressing
powered on. The oximeter will do self-test and then the software version and
warning message "Professional attendance is required for continuous monito-
ring!" will be shown on the screen, as shown in figure 4.1 (refer to your oximeter
for actual version).
34.7~37.3°C
35.5~37.5°C
36.6~38.0°C
35.8~38.0°C
Power/Left key for 1~2 seconds, then the oximeter will be
ENGLISH
Figure 4.1