Note: The oximeter is calibrated in the factory before sale, and there is no need
for user to calibrate again.

6 Over-limit Indication

6.1 Limit settings

• SpO
low limit setting range: 50%~99%.
2
• Pulse Rate limits setting range:
High: 100bpm--240bpm
During the measurement, if the measured value exceeds the preset value, the
alert beeping sound will be activated, the value that is over-limit will blink at the
same time.

6.2 Over-limit indication sound mute setting

• During the measurement, if the over-limit indication sound is set to on, short
time press Mute/Right key (
mute, but the over-limited value still keeps blinking. At this moment, the speaker
volume icon becomes
indication, and the speaker volume icon becomes
aker volume is set as 1 or 2 grid(s);
4 grids;
means the speaker volume is set as 5 or 6 grids.
• During the measurement, if the probe is off or disconnected, the message
"Check Probe" shows on the display screen. The alert sound starts (interval is 5
seconds) and lasts for about 3 minutes. If the probe is still off, the Oximeter will
power off automatically.

7 Packing List

1. An Oximeter
2. A SpO
probe
2
3. User Manual
4. A oximeter rubber cover
5. A charging base
6. A temperature probe (optional)
7. Charging cable (optional)
8. A USB data cable (optional)
Low: 30bpm--99bpm