Error Type B - Self-Test Failed - BIOTRONIK CardioMessenger Smart Manual Tecnico

Ocultar thumbs Ver también para CardioMessenger Smart:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Error Type B - Self-Test Failed

All symbols are flashing, indicating that the
CardioMessenger is not ready for use.
Repeat the self-test as the CardioMessenger was
not able to complete it.
1. If the CardioMessenger is connected to the
power supply brick, disconnect it.
2. Press the blue key on the right side of the
CardioMessenger for about two seconds.
3. Leave the CardioMessenger powered off for
about thirty seconds.
4. Power on the CardioMessenger by connecting
it to the power supply brick.
The CardioMessenger starts and automatically
repeats the self-test. When the self-test is
completed, the CardioMessenger is ready for
use.
The connection test can take up to 15 minutes.
If all symbols continue to flash, the
CardioMessenger is defective. Return it to your
physician.