The
clinician. If SmartStart is enabled,
breathe into the mask and will stop automatically when you take your mask off. This
means you do not have to press the Front key to begin or end treatment.
Note: Do not enable SmartStart when using a mask with an anti-asphyxia valve, such as
the Mirage Full Face Mask.
S
TOPPING
To stop treatment at any time, remove your mask and press the Front key
or
if your clinician has enabled the SmartStart function, simply remove your mask and
treatment will end.
U
SING THE
The
A
The mask-fitting feature delivers air pressure for a three-minute period, prior to starting
treatment, for checking and adjusting your mask fit to minimize leaks. To use the mask-
fitting feature:
1
Fit your mask as described in the mask user instructions.
2
Hold down the Front key for at least 3 seconds until air pressure delivery starts.
The following display will appear on the LCD screen indicating that the Mask-Fitting
feature is in operation. The flow generator will ramp to the Mask-Fit pressure and will
remain at this pressure for 3 minutes. A Mask-Fit star rating is also displayed. See
"Definitions of Mask-Fit Star Rating" on page 25.
Note: The Mask-Fit star rating display disappears after 3 minutes.
The Mask-Fitting feature can only be started from the Ramp or Settling screen.
In AutoSet mode, the mask-fit pressure is the highest pressure that you are likely to receive
based on the previous night's session or 10 cm H
In CPAP mode, the mask-fit pressure is the set treatment pressure or 10 cm H
whichever is greater.
24
™
S
S
has a function called SmartStart which can be enabled by your
UTO
ET
PIRIT
T
REATMENT
M
-F
F
ASK
ITTING
EATURE
S
S
mask-fitting feature can be used to help you fit your mask properly.
UTO
ET
PIRIT
Mask Fit
excelnt
A
S
S
will start automatically when you
UTO
ET
PIRIT
Mask-Fit star rating
*****
O, whichever is greater.
2
O,
2