13. Glass plate
Incorrect cleaning of the glass plate can result in water
drops forming on the glass plate.
The measurement result will be incorrect
!
The glass plate has a functional high-quality coating
that must not be destroyed.
!
Cleaning TTM 2104/2204 with a high-pressure cleaner,
hard and sharp objects or microfiber cloths damages
the glass plates, the coatings on the glass plates,
seals and cables.
i
The functional high-quality coating on the glass plate
cannot be reapplied.
If the coating is damaged, the module must be
replaced.
"
The glass plate must only be cleaned with a paper
towel moistened with clean water or the glass
squeegee supplied.
Robert Bosch GmbH
Glass plate | TTM 2104/2204 | 45
14. Graphical user interface
To identify the sources of errors more easily, it helps to
display certain data in the browser.
Link
Contents
/special/progress.xml
Update Status
/special/ttm_config.xml
Module configuration
/special/mode
Module mode
/special/user.xml
User settings
/special/permanent.log
Operating history of module
/cmd/system-info
Module status and system status
/special/last-xml-result
xml file
/cmd/take-picture
Current sensor image
14.1
Displaying the status
i
The master module displays information about itself
and the satellite.
The satellite module only displays information about
the master.
In the browser's address bar, enter the IP address of
the desired module, followed by
Fig. 32:
System information about master, information about both
modules
Fig. 33:
System information about satellite, information about
master only
en
en
.
/cmd/system-info
|