Dirty Field Calibration Body; Roi Tool Incorrectly Positioned - Bosch TTM 2104 Instrucciones De Servicio

Ocultar thumbs Ver también para TTM 2104:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

12.1.2
Dirt on the calibration body or glass plate 
Fig. 24:
Input image calibration body with object
Fig. 25:
Output image calibration body with missing pixels
The result of dirt on the calibration body and/or glass
plate is a faulty calibration.
"
Repeat the calibration after cleaning the calibration
body or the glass plate.
i
If the error persists despite cleaning, turn the ROI
tool. If the error also moves, replace the ROI tool. If
the error remains at the same location, look for the
source of the error in the module.
12.1.3
Extraneous light
Fig. 26:
Input image ROI tool with shadow
Fig. 27:
Output image ROI tool with missing pixels
Fig. 28:
Output striped image example with missing pixels
The result of extraneous light is a faulty calibration.
"
Place the system in the dark during the field calibration,
e.g. cover with a cloth.
Robert Bosch GmbH
Calibration (field calibration) | TTM 2104/2204 | 43
12.2

Dirty field calibration body

 
Fig. 29:
Input image ROI tool with object
Fine fibers remain on the surface of the calibration
body as the result of improper cleaning (with
disposable paper towel or cotton cloth).
 The result is a faulty calibration.
Wash the calibration body with water and detergent.
Dry with a linen cloth or allow to dry in the ambient air.
12.3

ROI tool incorrectly positioned

ROI tool placed in such a way as to not cover the entire
diagnostic opening.
Fig. 30:
Photo with diagonally shifted ROI determination tool
"
Place ROI tool parallel to the diagnostic opening so
that it is covered completely.
en
en
|
1 691 206 007
2017-06-19