Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
Power frequency (50/60Hz)
magnetic field
EN/IEC 61000-4-8
Note: U
is the a.c. mains voltage before applications of the test level.
T
The Model 4701 is suited for use in the electromagnetic environment specified below. The customer or the user of the
Model 4701 should assure that it is used in such an environment.
Immunity test
1-46
EMC Information
Guidance and manufacturer's declaration - electromagnetic immunity
<5% U
(>95% dip in U
T
for 0.5 cycle
40% U
(60% dip in U
T
T
5 cycles
70% U
(30% dip in U
T
T
25 cycles
<5% U
(>95% dip in U
T
for 5 sec.
3 A/m
Guidance and manufacturer's declaration - electromagnetic immunity
EN/IEC 60601 test level
4701-209-0 01 REV A
)
<5% U
(>95% dip in U
T
T
for 0.5 cycle
) for
40% U
(60% dip in U
T
T
5 cycles
) for
70% U
(30% dip in U
T
T
25 cycles
)
<5% U
(>95% dip in U
T
T
for 5 sec.
3 A/m
Compliance level
Main power quality should
)
T
be that of a typical
commercial or hospital
environment. If the user of
) for
the Model 4701 requires
continued operation during
) for
power main interruptions, it
is recommended that the
device be powered from an
)
T
uninterrupted power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Electromagnetic environment-
guidance
www.stryker.com