English
Specifications (Continued)
EN
Electrical
Circuit breaker, Thermal
Illumination
Battery voltage
Rated duty cycle
Class I Equipment: Equipment that protects against electrical shock and does not solely rely on basic insulation, but
which includes an additional safety precaution that is provided for the connection of the equipment to the protective earth
conductor in the fixed wiring of the installation that accessible metal parts cannot become live in the event of a failure of
basic insulation.
Environmental conditions
Ambient temperature
Relative humidity
(Non-Condensing)
Atmospheric pressure
Specifications listed are approximate and may vary slightly from product to product or by power supply fluctuations.
Stryker reserves the right to change specifications without notice.
1-8
Introduction
Standard 4 motor function: Head-Unit-Foot-Trend-Like
120VAC, 60 Hz, 10 Amp
Current leakage less than 300 microamperes (per UL 60601-1)
Hospital grade plug and 3-wire heavy duty cord
Compatible with non-flammable anesthetic agents and oxygen by
nasal catheter or mask type
Over voltage category — II (per IEC 60601-1)
120VAC - 1500V
250VAC, 10 Amp (0059-196-0 00)
Range: 100 lx to 1500 lx
9VDC — Nurse Call Option (Alkaline Battery)
3 minutes / 57 minutes — 3 minutes ON and 57 minutes OFF
Operation
104° F
(40° C)
50° F
(10° C)
75 %
30 %
1060 hPa
700 hPa
4701-209-0 01 REV A
Storage and transportation
104° F
(40° C)
-4° F
(-20° C)
75 %
10 %
1060 hPa
500 hPa
www.stryker.com