Operator's Manual
2. Rotate Light Source so that the bottom is accessible.
3. Unscrew the captive slotted screw and open the door.
4. Unscrew the captive slotted screw on the lamp latch.
5. Lift the latch and disconnect the cable.
6. Grasp the base of the lamp fixture and pull straight out.
7 .
Discard the old lamp fixture in the original packaging and in a safe manner to prevent
injury in the event of glass breakage.
8. Orient the "D" shape of the new lamp fixture with the internal chamber. Align the
rails of the fixture with the alignment notches.
9. Slide the fixture into the chamber and press in firmly until it "clicks" into place.
10. Re-connect the cable to the terminals and tighten the captive screw on the latch 1/4
turn past finger tight.
11. Close the door on the bottom of the Light Source and tighten the captive screw.
12. Rotate Light Source to upright position, re-attach plug, and turn power on.
Cleaning
CL Series Light Sources
The console may be cleaned using a clean soft cloth with the following cleaning agents:
•
Enzymatic Solution (Enzol)
•
Isopropyl Alcohol
•
Hydrogen Peroxide 3% USP
•
2% Glutaraldehyde Solution
•
Soap and Water
Note
Caution Do not splash cleaning agents into any openings, seams or electrical
components.
Turn off Light Source before cleaning.
Routinely inspect and clean all air intakes / out flows at the rear of the chassis for
lint or other debris.
Allow the console to completely dry before use.
Cleaning of any fiber bundles or accessories used with the CL Series light
sources should be per the manufacturer's instructions.
CL100 / CL300 Surgical Illuminator
11