Operating
+10 °C to +40 °C, ≤ 85 % RH
Temperature
Storage-
-5 °C to +50 °C, ≤ 85 % RH
Temperature
Power Supply
4 x 1,5 V AA batteries (alkaline type LR6)
Battery Life
For approx. 300 measurements, depending
on blood pressure level or inflation pressure
Accessories
With storage pouch, instruction for use, 4 pcs
"AA" battery
Protection class Internal power supply, IPX0, no AP or APG,
continuous operation
Classification
Application part Type BF
Device with internal power supply, protection
class II (in reference to the AC/DC mains part
and the device as an overall system)
Key to symbols Application part Type BF
Important!
Read the operating instruction
These specification are subject to change without notice for
purpose of improvement.

11. Mains part

Model No.
8194047/02
Input
100 V – 240 V ~, 50/60 Hz
Output
6 V, 600 mA, only in connection with beurer
blood pressure monitor
Supplier
Ansmann AG
Protection
This device is double isolated and protected
against short circuit and overload by a pri-
mary thermal fuse.
Make sure to take the batteries out of the
compartment before using the mains part.
Double isolated / equipment class 2
Enclosures
Equipment enclosed to protect against con-
and Protective
tact with live parts, and with parts which can
Covers
become live (finger, pin, hook test).
The operator shall not contact the patient
and the output plug of AC mains part simul-
taneously.
• This unit is in line with European Standard EN60601-1-2
and is subject to particular precautions with regard to elec-
tromagnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this instrument. More details can be requested from the
stated Customer Service address or found at the end of the
instructions for use.
• The instrument is in line with the EU Medical Products
Directive 93/42/EC, the German medical products act and
European standards EN1060-1 (non-invasive blood pres-
sure monitors, Part 1: General Requirements) and EN1060-3
(non-invasive blood pressure monitors, Part 3: Supplemen-
tary Requirements for Electromechanical Blood Pressure
Measurement Systems).