Please contact BUFFALO FILTER Customer Service to purchase the following accessories:
•
Replacement Filters
•
Electronic Footswitch
•
Remote Switch Activator
•
Hoses, Tubing, Laparoscopic Kits, Adapters, Wands & Other Accessories
1.3 Operational Information
The operational information contained in this section is intended for the customer review of regulatory
issues. The information pertains to the use of the products both domestically and internationally:
1. The BUFFALO FILTER PlumeSafe Smoke Evacuation System(s) complies with IEC601.1 electrical
specifications in the following systems:
100V-120V
220V-240V
2. The PlumeSafe Smoke Evacuation System(s) has a classification under the UL2601-1, Clause 5.1 of:
CLASS 1 Equipment
3. The PlumeSafe Smoke Evacuation System(s) has an identification under the UL2601-1, Clause 5.2
according to the degree of protection against electric shock:
Type CF Equipment
4. The PlumeSafe Smoke Evacuation System(s) has an identification under the UL2601-1, Clause 5.3
according to the degree of protection against harmful ingress of water:
Ordinary Equipment (enclosed equipment without protection against ingress of water).
5. The PlumeSafe Smoke Evacuation System(s) has a classification under the UL2601-1, Clause 5.4
according to the method of sterilization or disinfection recommended by BUFFALO FILTER:
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution, or soapy
water. Wipe dry with a clean cloth. Do not steam sterilize.
6. The PlumeSafe Smoke Evacuation System(s) has a classification under the UL2601-1, Clause 5.5
according to the degree of safety of application in the presence of flammable anesthetic mixture
with air or with oxygen or nitrous oxide:
Equipment not suitable for use in the presence of Flammable Anesthetic mixture with air or
with oxygen or nitrous oxide.
7. The PlumeSafe Smoke Evacuation System(s) has a classification under the UL2601-1, Clause 5.6
according to the mode of operation:
Continuous Operation
Upon request, BUFFALO FILTER will provide the following:
8.
Service and Repair Instructions, including Circuit Diagrams and Parts List
The fuses on the circuit board are to be serviced by an authorized BUFFALO FILTER technician
9.
as follows:
PPS604 100/120/220/240V, 50/60Hz use 0.03125 Amp 250 Volt fuse
The PlumeSafe Smoke Evacuation System(s) and all filters and accessories are not intended for contact with
patients.
BUFFALO FILTER®
Porta PlumeSafe® 604 Smoke Evacuation System – Rev C
50/60 Hz,
50/60 Hz
Page 4 of 14