Maintenance; General Maintenance Information; Cleaning; Periodic Inspection - Buffalo filter Porta PlumeSafe 604 Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

3.1 General Maintenance Information

This section contains information for ordinary upkeep of the BUFFALO FILTER PlumeSafe Smoke
Evacuation System. While the system has been designed and manufactured to high industry standards, it
is recommended that periodic inspection and performance testing be performed by a qualified
Biomedical Technician to ensure continued safe and effective operation.

3.2 Cleaning

The exterior of the system may be cleaned by wiping it with a cloth that has been dampened (not
dripping) with a mild detergent such as Windex® or Formula 409®. Windex® is a registered trademark
of the Drackett Products Company. Formula 409® is a registered trademark of the Clorox Corporation.

3.3 Periodic Inspection

The PlumeSafe Smoke Evacuation System should be visually inspected at least every year. This
inspection should include checks for:
Damage to the power cord.
Damage to the power plug.
Tightness of the power plug.
Proper mating, cleanliness and absence of damage to the filter inlet.
Obvious external or internal damage to the system.
3.4 Troubleshooting the System – see below.
PROBLEM
1. Smoke Evacuation
System is ON but suction is
minimal or none
2. Smoke Evacuation
System does not function
even though suction ON
button is depressed
BUFFALO FILTER®
Porta PlumeSafe® 604 Smoke Evacuation System – Rev C

Maintenance

Section 3.0
POTENTIAL CAUSE
1. Filter not seated
completely
2. Filter clogged
3. Vacuum hose clogged
4. Motor/blower obstructed
1. Not plugged into
electrical outlet
2. Fuses blown
3. Footswitch plugged in
4. Electronic system failure
5. Main power switch in
OFF position
6. Filter life expired or
invalid filter
Page 12 of 14

CORRECTIVE ACTION

1.
Re-install ViroSafe Filter, press firmly into
place
2.
Replace filter – use ViroSafe EY filter and
pre-filters for Erbium:YAG laser procedures
3.
Check or replace Smoke Evacuation Tubing.
4.
Call BioMed or BUFFALO FILTER
Technical Services at: 1-800-343-2324 or
(716) 835-7000
1. Check power outlet and connection to rear or
side panel of machine
2. Replace fuses located within the power entry
module.
3. Unplug footswitch
4./5.Check main power switch on rear or side of
system to ensure that it is in on the ON (|)
position
6. Call BioMed or BUFFALO FILTER
Technical Services at: 1-800-343-2324 or
(716) 835-7000
Tabla de contenido
loading

Tabla de contenido