14. Remove the circular protective cap. Prime
the infusion set until medication dispenses
from the tip. Ensure there are no air bubbles
in the tubing or AutoSoft 30. Consult your
pump manufacturer's instructions for use for
proper priming procedures.
15. Place a finger in front of the cannula housing
and insert the connector needle. Make
sure the connector locks with a "click". Fill
the empty space in the cannula with the
following prime:
U-100 INSULIN:
0.7 units (0.007 ml)
DISCONNECT
i.
AutoSoft 30 allows you to temporarily dis-
connect from your pump. Place a finger just
in front of the cannula housing and gently
squeeze the sides of AutoSoft 30. Pull the
connector needle straight out.
ii. Insert the circular protective cap into the
connector needle until you hear a "click".
iii. Insert the disconnect cover into the cannula
housing until you hear a "click".
RECONNECT
iv. Remove the circular protective cap. Prime
the infusion set until medication dispenses
from the needle.
When priming, hold AutoSoft 30
CAUTION:
with the needle pointing down.
v. Remove the disconnect cover from the
cannula housing.
vi. Place a finger in front of the cannula hous-
ing. Push the connector needle straight in
until you hear a "click".