SET-UP & OPERATION
1
2
Fully charge battery for 17
If applicable, ensure splash
HOURS. (7314P series only)
guard is securely attached to
inside of lid over filter
cartridge.
5A
5B
Disposable Container w/
Reusable Container w/external Bacteria Filter Connection:
Connect either end of the 4-3/8" tubing to the tubing connector
internal filter cartridge
then connect the other end to the bacteria filter. Ensure that the
Connection: Attach 4 ⁄ "
clear side of the bacteria filter is toward elbow and bottle when
tubing from filter cartridge to
installing. Do not reverse direction of filter. The bacteria filter
tubing connector on unit.
should then be connected to the 90˚ elbow connection, and the
elbow should be connected to the top of the container lid where
it says <Vacuum>.
NOTE–Inspect suction tubing and container for leaks, cracks, etc. and assure that all connections are secure and without leaks before using.
7A
7B
7314P - Select desired power
7314D series (non-battery
source. (Skip steps 8 if using
label)
internal battery power.)
NOTE– The 7314D series is not factory equipped with an
internal rechargeable battery. 7314P series is factory equipped
with an internal rechargeable battery and all information
regarding battery operation in this guide is applicable.
WARNING
12
If the unit is not receiving power from an external source or the battery was not recharged, the low battery indicator light will remain
on and the performance of the unit will drop off rapidly. Switch to another power source immediately after the low battery light appears
to avoid an interrupted suction procedure.
NOTE– Gauge is for reference only. If the unit sustains a severe drop, accuracy of the gauge must be checked.
CAUTION– When automatic float shut-off is activated, contents of the collection container should be emptied. Further suctioning could cause
damage to the vacuum pump.
CAUTION– Should fluid be aspirated back into the unit, equipment provider servicing is necessary as possible vacuum pump damage may
Verify suction level.
result.
NOTE - Always verify suction
level before beginning by
occluding open end of patient
tubing while observing gauge.
Adjust knob to desired level.
SE-7314-1
3
4
Securely attach lid to
Insert container into holder and gently push into place.
container.
NOTE – Do not use excessive force. Pushing container down
too hard could cause potential leak or loss of suction.
5C
6
Attach 6' patient tubing to
container lid at outlet labeled
<Patient>.
8
9
If using AC or DC power, plug
Plug the other end into an AC
the small connector into the
wall outlet or DC receptacle.
DC power input on the side of
NOTE– The AC adapter may
unit.
become warm to the touch
during charging or running of
the unit. This is normal.
7
Ensure power switch is
"off".
10
11
Adjust the suction level.
Turn the unit
"on".
EN - 5