TROUBLESHOOTING
DANGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a qualified
DeVilbiss Healthcare provider or an authorized service center. Opening or tampering with the unit will void warranty.
NOTE– Your DeVilbiss Suction Unit contains no user-serviceable parts. If you believe your unit is not working properly, BEFORE YOU RETURN IT TO YOUR DEVILBISS
HEALTHCARE PROVIDER WHERE YOU PURCHASED IT, PLEASE TAKE A FEW MOMENTS TO CHECK FOR THESE POSSIBLE CAUSES:
PROBLEM
Unit does not turn on when external power is connected. Green
external power light does not illuminate.
Pump runs, but there is no suction.
Low suction.
Unit does not turn on (no external power is connected). 7314P Series
Only
Battery will not charge (external power and charge indicator lights
should be illuminated during charge mode) 7314P Series Only
SPECIFICATIONS/CLASSIFICATIONS
Size (including container)
Weight (including container)
Typical Operating Sound Level
Electrical Requirements
Vacuum Range
Air Flow @ Pump Inlet
Disposable Collection Container Capacity
Reusable Collection Container Capacity
Warranty
Internal Battery (7314P Series Only)
Approvals
Adapter Manufacturer Information
Environmental Conditions
Operating Temperature Range
Operating Relative Humidity
Operating Atmospheric Pressure
Storage & Transport Temperature Range
Storage & Transport Relative Humidity
Storage & Transport Atmospheric Pressure
Equipment Classifications
With respect to protection from electric shock
Degree of protection against electric shock
Degree of protection against ingress of liquids
Mode of Operation
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide.
ISO Classification
7314P series only - Electrically powered medical suction equipment for field and transport use according to EN ISO 10079-1 : 2009
High Flow/High Vacuum
7314D series only - Electrically powered medical suction equipment for non-transport use according to EN ISO 10079-1 : 2009
* Conditions may vary based on altitude above sea level, barometric pressure, and temperature.
Manufacturer's Note
Thank you for choosing a DeVilbiss Suction Unit. We want you to be a satisfied customer. If you have any questions or comments, please send them to our address on the
back cover.
For Service Call Your Authorized DeVilbiss Healthcare Provider:
Phone
SE-7314-1
ACTION
1. Check power sources and connections.
2. Ensure wall outlet is live by plugging in a lamp.
1. Check that all tubing is connected properly.
2. Check tubing connections for breaks or leaks.
3. Ensure that float shut-off in collection container is not activated or filter cartridge occluded.
4. Check for leaks or cracks in collection container assembly.
1. Use vacuum regulator knob to increase suction level.
2. Check system for leaks.
1. Check that battery is fully charged and/or charge battery.
1. Verify that both external power and charge indicator lights illuminate.
2. Check power sources and connections.
3. Ensure wall outlet is live by plugging in a lamp.
8.3 H x 8.0 W x 8.5 D (21.1 cm x 20.3 cm x 21.6 cm) (not including AC to DC adapter)
7314P Series - 6.6 lb. (3 kg) (not including AC to DC adapter)
7314D Series - 4.3 lb. (2.0 kg)
55 dBA
100-240V~, 50/60Hz, 1.2A max
50-550 mm Hg +/- 10%*
27 LPM (free flow) typical (may be less when running from internal battery)* (NOTE– Using Flovac Liners may impair the device's
performance.)
800 ml (cc)
1200 ml (cc)
Two-years limited, excluding internal battery (7314P series only) and collection container
90-day
IEC 60101-1-2; CSA-C22.2 # 601.1; UL 60601-1; EN 60601-1-2; ISO 10079-1; IEC 60601-1; IEC 60529 IP12; IEC 60601-1-6;
CENELEC EN 60601-1
Emerson Model # AD5012N2LM or Autec Power Systems Model # DTM36-12 or SL Power/Ault Model # MENB1040A1240N02
32°F (0°c) - 104°F (40°c)
0-95%
10.2 psi (70 kPa) - 15.4 psi (106 kPa)
-40°F (-40°c) - 158°F (70°c)
0-95%
7.3 psi (50 kPa) - 15.4 psi (106 kPa)
Class II and internally powered
Type BF Applied Parts
IP12 and ordinary power supply
Intermittent Operation: 30 minutes on, 30 minutes off
Purchase Date
; 12V
; 33 W Max
Serial #
EN - 9