1.8.4. EMERGENCY BUTTON

The system is equipped with an emergency button to stop X-ray device operation, located under the patient support
arm in proximity of the telescopic column. A remote emergency button is connected using the dedicated connector on
the board located at the feet of the lifting column.
1 - Emergency button
These buttons must be activated in the event of danger and emergency, for example, failed interruption of radiation
from the source, in situations of evident danger to persons or when an emergency is signalled.
If the emergency button is pressed, radiation emission is immediately interrupted and all the movements of the patient
support and the rotary arm are stopped, all the linear motors run a distance of less than 10 mm and the rotary
movements a distance of less than 2 degrees, and the button remains locked in safe conditions.
After the emergency, turn the button in the direction indicated by the arrow to resume normal operation.

1.8.5. EXPOSURE TO LASER RADIATION

The system contains some class 1 LASER diodes, in compliance with IEC 60825-1.
Both the patient and the operator may be dazzled by laser traces.
-
Do not look directly into the laser beam. Pay attention to avoid that the laser beam hits the patient's eye.
-
Keep a distance of at least 10 cm between the eye and the laser.
The position of the laser sources is indicated with the following symbol