If the user would like to know the battery status of the Zn-Air Battery Pak they are currently
using and have an AB myPilot paired to their device, they can conduct a Battery Check readout
of the Naída CI. To ensure an accurate reading of the Zn-Air Battery Pak, wait at least 15
minutes after it has been attached to the Naída CI before performing a Battery Check readout.
Battery Check
Please see the
section of the
conduct a battery status check using AB myPilot.
Understanding Internal Alarms
Internal alarms are a programmable feature that can be used independently or in conjunction
with the Naída CI LEDs to provide important information about Naída CI status and battery life.
Naída CI Internal Alarms
Behavior
Programmable
Beeps upon
Yes
program change
Short Beep upon
Yes
increase/decrease
in volume
Long Beep (once
Yes
every 15 minutes)
Using the Naída CI Headpieces:
The Naída CI works with the Universal Headpiece (UHP)
The Universal Headpiece
The UHP is designed for use with the Naída CI processor during all daily activities, with the
exception of swimming and bathing.
Replacing RF Cables
RF cables are available in a variety of colors and lengths to suit the recipient's personal wearing
preferences. Before you can use the Naída CI sound processor, you must use the cable to
connect it to the headpiece. Inspect the cable on a regular basis to ensure it is in good condition.
26
Naída CI Sound Processor User Guide
AB myPilot User Guide
for instructions on how to
Indication
1 beep indicates program one
•
2 beeps indicate program two
•
3 beeps indicate program three
•
4 beeps indicate program four
•
5 beeps indicate program five
•
Beeps once per press of the volume control
either up or down (a double beep will be heard
when the following settings are reached: top of
the volume range; baseline volume setting; and
bottom of the volume range)
Low battery
If it appears damaged, replace it with a new cable.
To attach the cable to the headpiece:
1
Hold the headpiece in one hand and the
cable connector in the other hand.
To attach the cable to the sound processor:
1
Hold the sound processor in one hand and
the cable connector in the other hand.
For assistance in troubleshooting headpiece cables, please refer to the
of the User Guide.
Adjusting Headpiece Magnet Strength
It is important to have the correct magnet strength so the recipient does not experience
discomfort or retention issues. If an insufficient number of magnets are used in the headpiece,
it may fall off more often than is acceptable. If too many magnets are used , the recipient may
experience irritation or discomfort. Consult a hearing care professional if you have any concerns
regarding magnet strength. If deemed appropriate, your hearing care professional may insert
additional magnets or remove magnets from the headpiece. If the recipient experiences any
redness, irritation, or discomfort, discontinue use of the headpiece immediately and contact
Headpiece Instructions for Use
a cochlear implant center. See the
regarding adjustment of headpiece magnet strength.
2
Push the cable connector firmly into the
headpiece connector.
2
Push the cable connector firmly into the
sound processor connector port.
Troubleshooting
section
for additional information
27