MONITOR ONLY/Shutdown

6.3 MONITOR ONLY

(Patient Temperature Monitoring,
No Therapy)
Press MONITOR ONLY when no therapy is required,
but monitoring of patient temperature is desired.
To operate in MONITOR ONLY mode:
1.
Insert sensing end of patient probe into
patient and tape securely to prevent accidental
dislodgment.
2.
Insert plug end of patient probe into the
PATIENT PROBE jack. Make sure probe plug is
fully seated.
Use only Gaymar probes or equivalent YSI
400 series probes. If in doubt that the correct
probe is being used and is operating properly,
perform a PROBE CHECK test (p. 13).
3.
Press the MONITOR ONLY button. The upper
window on the button will light. The patient's
temperature will be displayed.
If the machine senses a patient probe temperature
below 29°C or above 40°C, the CHECK PATIENT
alert will light and an audible alarm will sound.
REMEMBER—
in MONITOR ONLY, no therapy is provided.
10

7.0 Shutdown Procedure

When therapy is completed, turn off the Medi-Therm II
machine. Close the pinch clamps before disconnecting the
blanket.
NOTE: Do not discard the DBK35CE Connector Hose.
Medi-Therm
TM
II