9.1.2 Parts Counting
Before the balance can count parts, it must know the average part weight (i.e. refer-
ence). This is either entered manually or determined by weighing. Proceed by putting
on a certain number of the parts to be counted. The balance determines the total
weight and divides it by the number of parts (the so-called reference quantity). Count-
ing is then carried out on the basis of the calculated average piece weight.
As a rule:
The higher the reference quantity the higher the counting exactness.
Display overview for parts counting operation:
1. Number of all parts on the weighing plate
2. Weighing assistance coarse/fine (only displayed for active dispensing)
3. Operating mode (status bar)
4. Reference weight
5. Weight of all parts on the weighing plate
6. Tolerance marker (only displayed for active tolerance weighing)
too light
Nominal
value
too heavy
7. Display for parts counting operation
Operation
Call up menu item "P1 Parts Counting" (see chap. 9.1)
Press
button
Use the cursor ( ► ) to select your setting (01 - 07)
01 Reference weight – manual entering of the reference weight
Press the
key; the current menu item blinks.
Use the arrow keys (see chapter 9.1) to enter the refer-
ence weight for a unit. Confirm by pressing the
58
Display
Parts counting setup
M1 01 Unit weight
02 Recall sample
03 Store sample
04 Checkweighting
05 Filling
06
Statistics
07 Run
key
ALTPLT-BA-e-0721
5.0000
g
ON
ON
OFF