8.3.5 P5 RS-232
In this menu item you can save your settings for the interface.
Operation
Call up menu item "P5 RS-232" (see chapt. 8.1)
Press
button
01 -10 Parameter selection
Use the cursor ( ► ) to select your parameters
(01 - 10)
01 Baud rate
02 Parity
03 Data bits
04 Stop bits
05 Handshake
06 Autom. Printout
38
2400
0
4800
1
9600
2
19200
3
none
0
even
1
uneven
2
7 Bit
1
8 Bit
2
1 Stop bit
1
2 Stop bit
2
None
0
RTS/CTS
1
XON/XOFF
2
None
0
steady
1
interval (dis-
2
crete)
at resting
3
Display
20.02.05
Setup
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
29.12.04
Setup
P 5 01 B ud rate
a
02 Parity
03 Data bits
04 Stop bits
05 Handshake
06 Auto print
07 Inetrval
08
Min.mass
09 Print on stab
No data output
Continuous serial data out-
put
Serial data output after a
certain interval, defined in
parameter „07 Interval"
Unique autom. Data output
for steady weighing value.
New output only after re-
moving the weight and new
loading
ALTPLT-BA-e-0721
9 6 0 0
2
0
no ne
2
8 bits
1
1 bit
no ne
0
no ne
0
2 5
* 0 .1 s
1
1 d
1 enabled