8.3.6 P6 Printouts
In this menu item you can select between 5 different types of data output:
Operation
Call up menu item "P6 Printouts "
(see chapt. 8.1)
Press
button
01 Printout no.
Use the cursor ( ► ) to select "01 Printout no."
Press the
key; the current menu item blinks
Use the
and
following settings:
0 standard
1 1. Printout
4 4. Printout
Use the
key to confirm your selection
8.3.6.1 Standard data output (0 standard)
Data output is carried out by actuating the
Definition of the log header is executed in menu item "P2 GLP".
Examples.
Date:
22/10/2004
Time:
13.04.23
User:
Mustermann
Project:
XW/456
0.008 g
40
keys to select between the
Date:
Time:
User:
Project:
Balance:
?
*: ? = instable measuring value
Display
20.02.05
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
29.12.04
P 6 01 Printout No.
02 Pr. 1 start
03 Pr. 1 stop
04
Pr. 2 start
05
Pr. 2 stop
06 Pr. 3 start
07 Pr. 3 stop
08 Pr. 4 start
09 Pr. 4 stop
10 String 1
11
String 2
88 String 79
89 String 80
key.
22/10/2004
13.16.49
Mustermann
XW/456
10
62.685 g *
ALTPLT-BA-e-0721
Setup
Setup
S ta n d a rd
0
0
0
0
0
0
0
0
0