KERN and SOHN ALT Manual De Instrucciones página 110

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Press the
key; the adjustment with internal ad-
justment weight is made.
After successful adjustment the balance automati-
cally returns to weighing mode.
In case of an adjustment error (e.g. objects on the
weighing plate) the display will show an error mes-
sage, repeat adjustment.
02 Ext. Cal adjustment with external weight - locked for verifiable devices
03 User cal. - locked for verifiable appliances
04 Cal. test
Here, deviation from the last adjustment is determined. This is only a check, i.e.
no values are changed.
Use the cursor ( ► ) to select "04 Cal. test"
Press
key; calibration test will be executed
automatically
The result is displayed
05 Weight corr. - Locked for verifiable appliances
06 Autom. cal. - locked for verifiable appliances
07 Auto. Cal.time – locked for verifiable appliances
08 Print report - printout of the adjustment data
Use the cursor ( ► ) to select "08 Print report"
Press
key, the current menu item blinks
With the arrow keys save your settings
Print report deactivated
0
Print report activated
1
Confirm by pressing the
ALT/PLT-BA-e-0721
key
Calibration
Internal calibration
27.01.05
Weighing
0.0000
0
0%
31.01.05
Setup
P 1 01 Int. calibr.
********
02 Ext. calibr.
********
03 User calibr.
********
04 Calibr. test
********
05 Weight. corr.
0 .0
06 Aut calibr.
o
0
07
Auto cal. time
1 1 h o u r
08
Print report
1
Calibration
Calibration check
Calibration
Check results
Cal.: 180.1354
Act.: 180.1438
Diff.: - 0.0084
31.01.05
Setup
P 1 01 Int. calibr.
********
02 Ext. calibr.
********
03 User calibr.
********
04 Calibr. test
********
05 Weight. corr.
0 .0
06 Aut calibr.
o
0
07
Auto cal. time
1 1 h o u r
1
08
Print report
g
100%
fun c ti o n
fun c ti o n
fun c ti o n
fun c ti o n
o ff
o n
fun c ti o n
fun c ti o n
fun c ti o n
fun c ti o n
o ff
o n