4.2 Remote control commands
80
IBM means the IBM character set following character set table 437 and IBM graphics
control characters. Select 'IBM' for work with the computer.
4.2.2.37. Config.PeriphUnit.RepToComport 1,2,1&2. And at 756: int.,1&int.,
2&int.,all. Selection of target for manually triggered reports.
int.
Internal printer.
1:
COM1
2:
COM2
4.2.2.38. Config.PeriphUnit.Balance
Selection of the balance type.
4.2.2.39. Config.PeriphUnit.Stirrer
Automatic stirrer control. With "ON" the stirrer will be switched on after starting of
conditioning. In the inactive state, the stirrer is switched off again.
4.2.2.40. Config.PeriphUnit.RemoteBox.Status
Config.PeriphUnit.RemoteBox.Keyboard
Config.PeriphUnit.RemoteBox.Barcode
Connections via Remote Box.
.Status:
Select if a Remote Box is connected.
.Keyboard: Type of keyboard which is connected to the Remote Box.
.Barcode: Select target in Coulometer where you wish to have the string from the
barcode reader. "input" means that the string comes into the field where
the cursor is currently placed.
4.2.2.41. Config.Aux.Language
Selection of the dialog language.
4.2.2.42. Config.Aux.Set
Config.Aux.Set.Date
Config.Aux.Set.Time
Date and time.
Input format of the date: Year-month-day, two-digit, enter leading zeros.
Input format for the time: Hours:minutes, two-digit, enter leading zeros.
Date and time have to be set with &Config.Aux.Set $G just after entry of the value.
4.2.2.43. Config.Aux.RunNo
Current sample number.
Set to 0 on power on and initialization. After 9999, counting starts again at 0.
Sartorius,Mettler,Mettler AT,
US, deutsch, francais,
español, schweiz.
input, method, id1, id2,
id3, smpl size
english, deutsch, francais, espanol,
italiano, portugese, svenska
YYYY-MM-DD
756/831 KF Coulometer, Instructions for Use
AND,Precisa
ON, OFF
ON, OFF
$G
HH:MM
0...9999