Metrohm 756 KF Instrucciones Para El Uso página 100

Coulometer
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

4.2 Remote control commands
92
Setup.SendMeas.Assembly.Pot
Setup.SendMeas.Assembly.IPulse
Setup.SendMeas.Assembly.Bur.V
Selection of the values from Assembly for the output in the set time interval (see
4.2.2.92):
.CyclNo:
Cycle number of the potential measurement. Together with the cycle
time (4.2.2.77), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and it is al-
ways incremented as long as the instrument remains switched on.
.I:
Total charge in mA·s associated to the cycle number, e.g. "667.48".
.Meas:
Measured value in mV associated to the cycle number, e.g. "104.2".
.Pot:
Voltage at generator electrode associated to the cycle number.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means
>28 V.
.IPulse:
Current of pulse associated to the cycle number.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
.Bur.V:
Dosed volume of connected Dosino in ml, e.g. "5.234".
The unit "assembly" must be preset (see 4.2.2.92).
4.2.2.94. Setup.SendMeas.Titrator.CyclNo
Setup.SendMeas.Titrator.Water
Setup.SendMeas.Titrator.Meas
Setup.SendMeas.Titrator.dWaterdt
Setup.SendMeas.Titrator.I
Setup.SendMeas.Titrator.Pot
Setup.SendMeas.Titrator.IPulse
Selection of the values from the titrator which are sent in the set time interval (see
4.2.2.91):
.CyclNo:
Cycle number. Together with the cycle time (4.2.2.78), a time frame can
be set up. The other data belong to the corresponding cycle number. The
cycle number is set to 0 at the start of a method and it is incremented
until the end of the method.
.Water:
Total water associated to the cycle number in ug, e.g. "62.313"
.Meas:
Measured value in mV at the indicator electrode associated to the cycle
number, e.g. "104.2".
.dWaterdt: Rate or drift associated to the cycle number in ug/min, e.g. "23.0".
.I:
Total charge in mA·s associated to the cycle number, e.g. "667.48".
.Pot:
Voltage at generator electrode associated to the cycle number.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3 means
>28 V.
.IPulse:
Current of actual pulse associated to the cycle number.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
OV will be sent for "overrange".
The unit "titrator" must be preset (see 4.2.2.92).
4.2.2.95. Setup.AutoInfo.Status
Setup.AutoInfo.P
Setup.AutoInfo.T.R
Setup.AutoInfo.T.G
Setup.AutoInfo.T.GC
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
ON, OFF
756/831 KF Coulometer, Instructions for Use
Tabla de contenido
loading

Este manual también es adecuado para:

831 kf5.756.00125.831.0011

Tabla de contenido