Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment specified be-
low. The customer or user of the transmitter should assure that it is used in such an
electromagnetic environment.
Immunity Test
Radiated RF
 IEC 61000-4-3
IEC 60601 Level
Compliance Lev-
3 V/m
10 V/m
 80 MHz to
 80 MHz to
2.5 GHz
6 GHz
Electromagnetic Environ-
el
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the transmitter, including ca-
bles, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmit-
ter.
Refer to the recommended
separation distance table for
more information.
d = 0.35 P
 80 MHz to 800 MHz
d = 0.70 P
 800 MHz to 6 GHz
Where P is the maximum out-
put power rating of the trans-
mitter in watts (W) according to
the transmitter manufacturer
and d is the recommended
separation distance in meters
(m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site sur-
a
vey
, should be less than the
compliance level in each fre-
quency range
Interference may occur in the
vicinity of equipment marked
with the following symbol:
-26-
ment Guidance
b
.