Guidance and Manufacturer's Declaration - Electromagnetic Immunity

The transmitter is intended for use in the electromagnetic environment specified be-
low. The customer or the user of the transmitter should assure that it is used in such
an environment.
Immunity Test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transi-
ent/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short in-
terruptions and voltage
variations on power
supply lines
IEC 61000-4-11
Power frequency
 (50/60 Hz) magnetic
field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 Test Level
±2 kV, ±4 kV, ±8 kV, ±15 kV
Air
±2 kV, ±4 kV, ±6 kV, ±8 kV
Contact
±2 kV for power supply
lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5% U
(>95% dip in U
T
T
for 0.5 cycle
3 A/m
Compliance
Level
±2 kV, ±4 kV,
±8 kV, ±15 kV
Air
±2 kV, ±4 kV,
±6 kV, ±8 kV
Contact
Not applicable
Not applicable
)
Not applicable
400 A/m
-24-
Electromagnetic Envi-
ronment - Guidance
For use in a typical do-
mestic, commercial, or
hospital environment.
Requirement does not
apply to this battery pow-
ered device.
Requirement does not
apply to this battery pow-
ered device.
Requirement does not
apply to this battery pow-
ered device.
Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typi-
cal domestic, commer-
cial, or hospital
environment.