Information On The Power Supply/Battery Charging - Ottobock Axon-Bus Prosthetic System Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 83
CAUTION
Unintentional unlocking of the Axon-Bus gripping component
Injuries due to releasing the Axon-Bus gripping component from the forearm (e.g. while carrying
objects).
The two release buttons should only be activated in order to change the Axon-Bus gripping
component, deliberately and with due consideration of the respective situation.
NOTICE
Improper care
Damage or fracture due to embrittlement of the plastic through the use of acetone, petrol or sim­
ilar solvents.
Clean the Axon-Bus prosthethic system with a damp cloth and mild soap only.
Use only the following products for cleaning/disinfecting the inner socket:
Cleaning: 453H10=1 Ottobock DermaClean
Disinfection: 453H16 Ottobock DermaDesinfect

4.6 Information on the Power Supply/Battery Charging

CAUTION
Dirtying of the electrical contacts
Injuries due to faulty control or malfunction of the Axon-Bus prosthethic system caused by an
insufficient supply of power to the Axon-Bus components.
Ensure that the contacts of the charging receptacle are always clean and free of grease.
Clean the electrical contacts of the battery charger and charging receptacle regularly using
cotton swabs and a mild soap solution.
Avoid damaging the contact surface coating with pointed or sharp objects.
NOTICE
Use of incorrect power supply unit
Destruction of Axon-Bus prosthesis system due to incorrect voltage, current or polarity.
To charge the Axon-Bus prosthesis system use only the supplied 757L500 AxonCharge
Integral power supply unit.
NOTICE
Contact of the charging plug with magnetic data carriers
Loss of data or wiping of the data carrier.
Do not place the magnetic charging plug on data carriers such as credit cards, diskettes,
audio and video cassettes or the like.
INFORMATION
The Axon-Bus prosthethic system cannot be used as long as the charging plug is connected to
the charging receptacle. The Axon-Bus prosthethic system is deactivated for the duration of
charging.
Ottobock | 29