ENGLISH
A: 
Schuko socket 
230VAC ±10% 
B: 
¾ G connection 
C: 
ref. CM61031, length 3 metres with ¾ G 
connection 
D: 
Gravity discharge 
ref. CM61032, length 0.8 ‐ 2.7 metres 
E: 
Power cable  
L = 2.5 m 
Ref. CE06006 
Pct. 4
 
The disinfectant/detergent tank has a capacity of about 2 litres. 
Open the tank cover and remove the plug on the top of the tank (Pct. 5).  
Pour the liquid disinfectant/detergent into the tank 
(preferably using a funnel) until an acoustic/visual signal 
(visible on the display) indicates that the tank is full. 
Use only products recommended by the 
manufacturer. 
Do not use acid solutions or strong alkaline solutions such as sodium hypochlorite. 
 
IT'S NECESSARY TO EMPTY THE TANK BEFORE INSERT 
ANOTHER TYPE OF LIQUID.  
 
IT IS ABSOLUTELY FORBIDDEN MIX THE DIFFERENT LIQUID IN 
THE TANK 
 
Emptying the tank: 
When the device is not doing the cycle, press and hold the button 
subsequently push and keep pushed the button 
After 5 seconds will start the peristaltic pump, and an intermittent beep will confirm 
that it is running. When the tank is empty the pump will turn off automatically. 
At anytime you can interrupt the process by pushing the button 
B
Filling the disinfectant/detergent tank: 
A
E
C
D
14
Pct. 5
 and then 
 
 
loading